  • Measure DC and AC voltage and current, resistance, frequency, and period with laboratory accuracy and incredible system speed. Delivers results and convenient features that minimize test and troubleshooting times. One-Year Warranty.(USA)

  • Specifications:

    DC voltage
    Measurement method: Continuously integrating multi-slope III A-D converter
    A-D linearity: 0.0002% of reading + 0.0001% of range
    Input resistance:
    10 MΩ or 0.1 V, 1 V, 10 V ranges: Selectable > 10,000 MΩ
    100 V, 1000 V ranges: 10 MΩ ± 1%
    Input bias current: < 30 pA at 25°C
    Input protection: 1000 V all ranges
    dcV:dcV ratio accuracy:
    Vinput Accuracy + Vreference Accuracy

    True rms AC voltage
    Measurement method:
    ac coupled true rms - measures the ac component of the input with up to 400 Vdc of bias on any range
    Crest factor: Maximum of 5:1 at full scale
    Additional crest factor errors (non-sinewave):
    Crest factor 1-2: 0.05 % of reading
    Crest factor 2-3: 0.15 % of reading
    Crest factor 3-4: 0.30 % of reading
    Crest factor 4-5: 0.40 % of reading
    Input impedance: 1 M. ± 2% in parallel with 100 pF
    Input protection: 750 Vrms all ranges

    Resistance

    Measurement method:
    Selectable 4-wire or 2-wire Ohms.
    Current source referenced to LO input.
    Maximum lead resistance (4-wire): 10% of range per lead for 100 Ω and 1 kΩ ranges. 1 kΩ per lead on all other ranges.
    Input protection: 1000 V all ranges

    DC current

    Shunt resistance: 5 Ω for 10 mA,100 mA; 0.1 Ω for 1 A, 3 A
    Input protection:
    Externally accessible 3 A 250 V fuse
    Internal 7 A 250 V fuse

    True rms AC currentN

    Measurement method:
    Direct coupled to the fuse and shunt.
    ac coupled true rms measurement (measures the ac component only).
    Shunt resistance: 0.1 Ω for 1 A and 3 A ranges
    Input protection:
    ExternalIy accessible 3 A 250 V fuse
    Internal 7 A 250 V fuse

    Frequency and period

    Measurement method: Reciprocal counting technique
    Voltage ranges: Same as ac voltage function
    Gate time: 1 s, 100 ms, or 10 ms

    Continuity/diode

    Response time: 300 samples/s with audible tone
    Continuity threshold: Selectable from 1 Ω to 1000 Ω

    Measurement noise rejection 60 (50) Hz1

    dc CMRR: 140 dB
    ac CMRR: 70 dB
    Integration time and normal mode rejection2
    100 plc/1.67 s (2 s): 60 dB3
    10 plc/167 ms (200 ms): 60 dB3
    1 plc/16.7 ms (20 ms): 60 dB
    < 1 plc/3 ms or 800 µs: 0 dB

    Operating Characteristics4
    System speeds6
    Configuration rates: 26/s to 50/s
    Autorange rate (dc Volts): > 30/s
    ASCII readings to RS-232: 55/s
    ASCII readings to GPIB: 1000/s
    Maximum internal trig. rate: 1000/s
    Max.ext.trig.rate to memory: 1000/s
    Triggering and memory
    Reading HOLD sensitivity: 10%, 1%, 0.1%,or 0.01% of range
    Samples/trigger: 1 to 50,000
    Trigger delay: 0 to 3600 s: 10 µs step size
    External trigger delay: < 1 ms
    External trigger jitter: < 500 µs
    Memory: 512 readings
    Math functions NULL, min/max/average, dBm, dB, limit test (with TTL output)
    Standard programming languages: SCPI (IEEE-488.2), Agilent 3478A, Fluke 8840A/42A

    Accessories included: Test lead kit with probe, alligator, and grabber attachments Operating manual, service manual, test report, and power cord

    General specifications
    Power supply: 100 V/120 V/220 V/240 V ±10%
    Power line frequency:
    45 Hz to 66 Hz and 360 Hz to 440 Hz
    Automatically sensed at power-on
    Power consumption: 25 VA peak (10 W average)
    Operating environment:
    Temperature: RTD, Thermistor
    Full accuracy for 0°C to 55°C
    Full accuracy to 80% R.H. at 40°C Storage environment: - 40° C to 70°C
    Weight: 3.6 kg (8.0 Ibs)
    Safety: Designed to CSA, UL-1244, IEC-348 RFI and ESD: MIL-461C, FTZ 1046, FCC
    Vibration and shock: MIL-T-28800E, Type III, Class 5 (sine only)
    Warranty: 1 year

    Notes:
    1. For 1 kΩ unbalance in LO lead.
    2. For power line frequency ± 0.1%.
    3. For power line frequency ± 1% use 40 dB or ± 3% use 30 dB.
    4. Reading speeds for 60 Hz and (50 Hz) operation.
    5. Maximum useful limit with default settling delays defeated.