  • Measure DC and AC voltage and current, resistance, frequency, and period with laboratory accuracy and incredible system speed. Delivers results and convenient features that minimize test and troubleshooting times. One-Year Warranty.(USA)

    Features:

    DMM 6-1/2 Digit Dual Display, Enhanced Performance
    50,000 readings/sec. @ 4 1/2 digits to the PC
    1M reading memory
    Analog level and pre/post triggering
    Null, statistics, dBm, dB, Limit Test
    LAN, USB & GPIB standard
    Capacitance & temperature measurements
    Data logger for improved usability
    50,000 reading non-volatile memory
    LXI Class C Compliant

    Warranty: 1 year

    Notes:
    1. For 1 kΩ unbalanced in LO lead, ± 500 V peak maximum
    2. For 1 kΩ unbalanced in LO lead and <60 Hz, ± 500 V peak maximum
    3. Maximum rate for DCV, DCI, and resistance functions (using zero settling delay, autozero off, manual range)
  • Specifications:

    DC Voltage
    Measurement Method:

    Continuously integrating multi-slope IV A/D converter
    Linearity: 0.0002% of reading
    (10 V range) + 0.0001% of range
    Input Resistance:
    0.1 V, 1 V, 10 V: 10 Ω or > 10 GΩ
    Ranges: (Selectable)
    100 V, 1000 V: 10 MΩ ±1%
    Ranges: (Fixed)
    Input Bias Current: < 50 pA at 25°C
    Input Protection: 1000 V
    DC CMRR: 140 dB1

    True RMS AC Voltage
    Measurement Method:

    AC-coupled True RMS measurement.
    Digital sampling with anti-alias filter.
    Crest Factor:
    No additional error for crest factors < 10.
    Limited by peak input and 300 kHz bandwidth.
    Peak Input: 300% of range or 1100 V
    Overload Ranging: Will select higher range if peak input overload is detected during auto range. Overload is reported in manual ranging.
    AC CMR: 70 dB2
    Maximum Input: 400 Vdc, 1100 Vpk
    Input Impedance: 1 Ω ± 2% in parallel with < 150 pF
    Input Protection: 750 Vrms all ranges

    Resistance Measurement Method:
    Selectable 2-wire or 4-wire.
    Current source referenced to LO input.
    Offset Compensation: Selectable on the 100 Ω, 1 kΩ, and 10 kΩ ranges
    Max. Lead Resistance (4-wire): 10% of range per lead for 100 Ω, 1 kΩ 1 kΩ per lead on all other ranges
    Input Protection: 1000 V on all ranges

    DC Current
    Current Shunt:

    200 Ω for 100 µA, 1 mA
    2 Ω for 10 mA, 100 mA
    0.1 Ω for 1 A, 3 A
    Input Protection: 3 A, 250 V fuse

    True RMS AC Current
    Measurement Method:
    AC-coupled True RMS measurement.
    Directly coupled to the fuse and shunt.
    Digital sampling with anti-alias filter.
    Current Shunt:
    200 Ω for 100 µA,
    1 mA 2 Ω for 10 mA, 100 mA
    0.1 Ω for 1 A, 3 A
    Maximum Input:
    The peak value of the DC + AC current must be <300% of range. The RMS current must be <3 A including the DC current content.
    Input Protection: 3 A, 250 V fuse

    Frequency and Period

    Measurement Method:
    Reciprocal-counting technique. AC-coupled input using the AC voltage measurement function.
    Input Impedance:
    1 M&Omdega; ± 2% in parallel with < 150 pF
    Input Protection: 750 Vrms all ranges

    Capacitance
    Measurement Method: Current input with measurement of resulting ramp.
    Connection Type: 2-wire

    Temperature
    Thermistor: 2.2 kΩ, 5 kΩ, and 10 kΩ
    RTD: a = 0.00385 Ro from 49 Ω to 2.1 kΩ

    Continuity/Diode Test
    Response Time:

    300 samples/sec with audible tone
    Continuity Threshold: Fixed at 10 Ω

    Operating Characteristics

    Maximum readings/second
    Function3: 4.5,, 5.5, 6.5 Digits
    DCV: 50 k4 10 k, 1 k
    2-wire Ω: 50 k4, 10 k, 1 k
    DCI: 50 k4, 10 k, 1 k

    Frequency: 500, 90, 10
    Period: 500, 90, 10
    Filter setting: fast, med, slow
    ACV: 500, 150, 50
    ACI: 500, 150, 50


    Additional 34411A Specifications:
    Overall Bandwidth, DCV & DCI:
    15 kHz typical @ 20 µs aperture (-3 dB)
    Triggering: Pre/Post, Int/Ext, Pos/Neg
    Timebase Resolution: 19.9524 &micros 0.01% accuracy
    Trigger Jitter: 2 µs (p-p), 20 µs (p-p) when pre-triggered

    Triggering and Memory
    Reading Hold Sensitivity: 1% of reading
    Samples per Trigger: 1 to 1,000,000
    Trigger Delay: 0 to 3600 s (20 µs step size)
    External Trigger: Programmable edge, Low-power TTL compatible
    Delay: < 1 µs Max rate: 5,000/s
    Jitter: < 1 µs Min Pulsewidth: 1 µs
    Voltmeter Complete: 3 V Logic output, 2 µs pulse with programmable edge
    Nonvolatile Memory: 50,000 readings
    Volatile Memory: 1,000,000 readings
    Sample Timer:
    Range:
    0 to 3600 s (20 µs step sizes)
    Jitter: < 100 ns

    General Specifications:
    Power Supply: 100 V/120 V/220 V/240 V ±10%
    Power Line Frequency: 45 Hz to 66 Hz and 360 Hz to 440 Hz, Automatically sensed at power-on
    Power Consumption: 25 VA peak (16 W average)
    Operating Environment: Full accuracy for 0°C to 55°C, 80% R.H. at 40°C non-condensing
    Storage Temperature: -40°C to 70°C
    Weight: 3.72 kg (8.2 lbs)
    Safety: IEC 61010-1, EN 61010-1, UL 61010-1, CAN/CSA-C22.2 No. 61010-1, Refer to Declarations of Conformity for current revisions. Measurement CAT II 300 V, CAT I 1000 V. Pollution Degree 2
    EMC: IEC 61326, EN 61326, CISPR 11, ICES-001, AS/NZS 2064.1, Refer to Declaration of Conformity for current revisions.
    Vibration & Shock: MIL-T-28800E, Type III, Class 5 (Sine Only)
    LXI Compliance: LXI Class C, ver. 1.0